How to change the wand cable – iTero Lumina™
1. Power off the scanner and unplug the wand cable from behind the screen - refer to the location of the cable here.
2. Remove the wand from the cradle.
3. Use the Allen Key provided in the replacement kit to unscrew the cable.
4. Remove the cable straight out.
5. Insert the cable back ensuring correct alignment.
6. Use the Allen Key to tighten.
7. Place the wand back in the cradle and connect the cable to the back of the screen.
8. Power on the scanner and test scan to ensure the scanner is working.